Feature Summary
No need for costly extra sensors
A compact 3.4 cm diameter water level sensor that fits into tight spaces, the HYDROS 21 CTD sensor is a low-cost, durable, and easy-to-use tool for monitoring EC, temperature, and depth in both groundwater and surface water. More importantly, it’s an all-in-one instrument. When used with the ZL6 data logger or any other logging system, you won’t require an additional sensor for measuring—and referencing—barometric pressure. With a range of 0 to 120 dS/m and a measuring depth of 10 m, the sensor makes accurate depth and EC measurements for a broad range of applications.
3.4 cm
diameter sensor
Water level sensor with easy, research-grade data in real time
One of the qualities that makes the HYDROS 21 water level sensor so easy to use is its fast plug-and-play compatibility with the ZL6 data logger (also compatible with Campbell Scientific data loggers). Plus, with the ZL6, you can monitor and remotely access cloud data wirelessly from any internet-connected computer or device in near-real time.
Affordable, accurate, and reliable
While engineering the HYDROS 21 water level sensor, we wanted to design a water level sensor that was not only highly accurate but also provided a high degree of value. Because of its low-complexity, low-maintenance, and low-cost design, you can be assured you’re getting an instrument that will save you time, hassle, and money.
Best price-to-performance ratio on the market
Trying to measure water level, temperature, and EC typically requires an expensive CTD sensor with integrated telemetry. Unlike competitor sensors, all of the HYDROS 21’s data management happens in an external data logger, so the cost of each individual sensor is lower. Now, instead of having to rely on a single point measurement, you can take readings in several different locations without exceeding your budget.
TECHNICAL SPECIFICATIONS
Measurement Specifications
Water Depth:
Range: 0 – 10,000 mm
Resolution: 1 mm
Accuracy: ±0.25% of full scale at 20 °C
NOTE: Depth measurement accuracy assumes no abrupt temperature variations.
Temperature:
Range: -40 to +60 °C
Resolution: 0.10 °C
Accuracy: ±1.00 °C
NOTE: The pressure transducer may be damaged or destroyed if frozen in ice. Remove the sensor if the water temperature could drop below 0°C (32 °F).
Bulk Electrical Conductivity (EC):
Range: 0 – 120 dS/m
Resolution: 0.001 dS/m
Accuracy: ±0.01 dS/m or ±10%, whichever is greater
NOTE: The EC measurement is corrected to a standard temperature of 25°C.
Communication Specifications:
Output
DDI serial or SDI-12 communications protocol
Data Logger Compatibility:
METER ZL6 data loggers and any data acquisition system capable of 4.0- to 15.0-VDC power and serial or SDI-12 communications
Physical Specifications:
Dimensions
Diameter: 3.4 cm (1.3 in)
Length: 9.0 cm (3.5 in)
Operating Temperature Range
Minimum: 0 °C
Maximum: 60 °C
NOTE: Sensors may be used at higher temperatures under certain conditions; contact Customer Support for assistance.
Cable Length
10 m (standard)
20 m
40 m (maximum cable length)
Cable Diameter
6.00 ±0.25 mm (0.240 ±0.010 in) with minimum jacket of 0.80 mm (0.031 in)
Connector Types
Stereo plug connector or stripped and tinned wires
Stereo Plug Connector Diameter
3.5 mm
Conductor Gauge
20-AWG signal and power / 21-AWG ground wire
Electrical and Timing Characteristics
Supply Voltage Power to Ground
Minimum: 4.0 V
Typical: NA
Maximum: 15.0 V
Digital Input Voltage (Logic High)
Minimum: 2.8 V
Typical: 3.6 V
Maximum: 5.0 V
Digital Input Voltage (Logic Low)
Minimum: -0.3 V
Typical: 0.0 V
Maximum: 0.8 V
Digital Output Voltage (Logic Low)
Minimum: NA
Typical: 3.6 V
Maximum: NA
Power Line Slew Rate
Minimum: 1.0 V/ms
Typical: NA
Maximum: NA
Current Drain (During Measurement)
Minimum: 2.5 mA
Typical: 3.5 mA
Maximum: 8.0 mA
Current Drain (While Asleep)
Minimum: NA
Typical: 0.35 mA
Maximum: NA
Power Up Time (DDI Serial)
Minimum: 52 ms
Typical: NA
Maximum: 100 ms
Power Up Time (SDI-12)
Minimum: NA
Typical: 700 ms
Maximum: 750 ms
Power Up Time (SDI-12, DDI Serial Disabled)
Minimum: NA
Typical: 175 ms
Maximum: 200 ms
Measurement Duration
Minimum: NA
Typical: 450 ms
Maximum: 600 ms
Other
Compliance
EM ISO/IEC 17050:2010 (CE Mark)
Prop 65 warning
GSA
View GSA details
Your review appreciation cannot be sent
Report comment
Report sent
Your report cannot be sent
Write your review
Review sent
Your review cannot be sent
Reference: H21-USB
Brand: ONSET
Reference: 28010011
Brand: FUEL CELL
Reference: SONIC15
Brand: MAE
Reference: U30-NRC-000-10-S100-000
Brand: ONSET
Reference: 70010001
Brand: FUEL CELL
Reference: 28010001
Brand: FUEL CELL
Reference: SFL
Reference: CRONOSONIC
Brand: MAE
Reference: 28010017
Brand: FUEL CELL
Reference: 28010006
Brand: FUEL CELL
Reference: ZL6 BASIC
Brand: METER GROUP
Reference: U30-NRC-SYS-C
Brand: ONSET
Reference: RX3004+SP-611
Brand: ONSET
Reference: RK500-13
Brand: RIKA SENSORS
Reference: RX2101+SP-611
Brand: ONSET
Feature Summary
check_circle
check_circle