TCOMSYS01 Hot Cube Thermal Comfort Measurement System
TCOMSYS01 is a measuring system to help understand and quantify “causes and effect” leading to human thermal comfort. The TCOM01 body is temperature stabilised at 33 °C, so that it offers a relatively direct measurement of the human experience. In essence, TCOM01 is a miniature thermal mannequin, measuring according to the innovative Hot Cube method. The heater power required to keep the TCOM01 at a constant temperature is the main measurand. Incorporating 5 heat flux sensors with a black absorber, it also offers a detailed picture of the heat gain and loss from different directions, and a good indication of convective and radiative asymmetry. Other measurements are sensor body temperature, air temperature and relative humidity. In its standard configuration, the system consists of an MCU (Measurement and Control Unit) and a TCOM01 sensor. The MCU offers direct connection to any local area network and “Ethernet over USB”. TCOM01 is also available as a sensor only.
Measurand | heating power for stabilisation, heat flux (5x), ambient air temperature, relative humidity, TCOM01 body temperature |
TCOMSYS01 specifications | : |
TCOM01 temperature setting | 33 °C (user adjustable) |
Included cable | 2 x cable with 2 connectors (1.5 m) |
Standard rated operating temperature range | 10 to +25 °C |
Limiting temperature range | 25 to +50 °C |
Required data analysis | to be performed by the user |
Data display | in web browser |
Start up interval | < 10 min |
TCOM01 specifications | : |
Weight | 1.12 kg |
Temperature sensor | 10 kΩ thermistor |
IP protection class | IP65 |
Mounting | tripod with ¼ inch -20 UNC screw; a Manfrotto Pixi EVO 2-section tripod is included |
MCU specifications | : |
User interface | on MCU as a web page |
Connection | via LAN or "Ethernet over USB" |
Voltage measurement accuracy | 4 x 10-6 V (0.8 W/m²) |
Rated power supply voltage | 10 to 16 VDC |
Adapter power supply | 100 – 240 VAC, 50/60 Hz |
Rated operating temperature range MCU | 25 to +50 °C |
Data storage capacity | 2GB; > 1 year of data |
Introduction
TCOMSYS01 was originally designed to study the effect of radiation sources on human comfort. Equipped with heat flux sensors that measure in 5 directions and a humidity and temperature probe, it offers a good picture of energy gains and losses from all sides to a metal body that thermally resembles the human body; the TCOM01 sensor is in essence a miniature thermal mannequin.
TCOMSYS01 Hot Cube Thermal comfort measuring system
The TCOMSYS01 system employs dedicated sensors and electronics, measuring thermal comfort according to the new Hot Cube method. The high accuracy of the MCU (Measurement and Control Unit) ensures that TCOMSYS01 will still measure down to very low heat fluxes. The MCU has a robust aluminum housing. The system generates a measurement file, including a time-stamp. The measurement data are stored in the MCU and are later downloaded to a PC. The user is responsible for data analysis.
TCOMSYS01 Hot Cube Thermal comfort measuring system
How to use TCOMSYS01
The primary source of information from TCOMSYS01 simply is the power [W] required to keep the TC0M01 sensor at a constant temperature; a very direct measure of human comfort. This power may be compared to the power required at 20 °C ambient air temperature, no convection (zero wind speed) and no radiation. Power consumption will immediately show if there is a situation of overheating or heatstress or a situation of overcooling or coldstress.
The second direct information supplied by the heat flux sensors of TCOMSYS01 is the heat loss or gain [W/m²] as a function of direction. If radiative sources are dominant, TCOMSYS01 will measure radiative asymmetry.
To study the exact effect of radiative sources, in many cases a simple experiment, like switching a radiation source [on] and [off] or shielding a radiation source, can be used to distinguish between convective and radiative heat transport.
Far-infra red radiation sources may be studied by temporary shielding with silicon wafer. Silicon has around 50% transmission. Radiation from the sun or lamps may be quantified by shading and unshading.
To study the insulating effect of clothing, TCOM01 may be insulated using the same fabrics.
A web browser may be used for real time measurement review, data collection and for changing control settings such as TCOM01 body temperature.
What makes TCOMSYS01 different?
The main contributors to thermal comfort are air temperature, air speed, radiant temperature and humidity. Apart from this, there are personal contributors such as metabolic rate and insulation by clothing.
Many studies use Fanger’s thermal comfort equation as applied in EN ISO 7730: Moderate Thermal Environments - Determination of the PMV and PPD indices and specification of the condition for thermal comfort. Another approach utilises a globe temperature measurement: EN ISO 27243: Hot environments. Estimation of the heat stress on working man, based on the WBGT Index (Wet Bulb Globe Temperature).
These methods are quite indirect in particular when determining the heat flow from air speed and radiation.
TCOMSYS01 offers heat flux measurements. This approach is a lot more direct than the indirect estimate from air speed and radiation.
TCOMSYS01 offers directional information.
TCOMSYS01 works at a realistic skin temperature of 33 °C (user adjustable).
Rated operating conditions
TCOMSYS01 is designed to work between +10 and +25 °C. In its standard configuration, uninsulated and stabilised at 33 °C, it is rated for operation in indoor conditions including significant radiative heating.
It will stabilise at 33 °C under approximately one of the following conditions:
Under other conditions, the sensor may not be able to stabilise its body temperature due to overheating (indicating overheating or heatstress) or shortage of electrical power (indication of overcooling or coldstress). Powered using a low voltage, TCOMSYS01 is safe to use.
TCOMSYS01 may be used for short-term outdoor experiments in the order of weeks. Under long-term exposure to solar radiation, the black heat flux sensor coating may become more reflective.
TCOMSYS01 Hot Cube Thermal comfort measuring system
User interface: MCU is a web server
The MCU serves as a web server, and can be connected to any local area network. No more downloading of USB drivers and special interface software! Alternatively it offers an “Ethernet over USB” or virtual Ethernet link, where you connect to the MCU using a USB cable. If you type into your web browser the MCU’s IP address (192.168.66.1 by default), you have access to the user interface.
TCOMSYS01 Hot Cube Thermal comfort measuring system
Ordering the TCOM01 sensor only
The sensor TCOM01 is also available as a “sensor only”; The configuration then includes the mannequin with 5 x heat flux, 1 x temperature, 1 x heater, 2 x cable, 2 x chassis connector and 1 x tripod. The user then must combine it with his or her own measurement and control unit.
Suggested use
Your review appreciation cannot be sent
Report comment
Report sent
Your report cannot be sent
Write your review
Review sent
Your review cannot be sent
Reference: FS-1000
Brand: BOSEAN
Reference: HOBOnet T12
Brand: ONSET
Reference: UG-5-AFR-24V
Brand: FUEHLER SYSTEME
Reference: DV-DENDRO
Brand: ECOMATIK
Reference: U24-002-C
Brand: ONSET
Reference: RK400-13
Brand: RIKA SENSORS
Reference: Evolution-M
Brand: NESA
Reference: MQ21
Brand: Meatrol
Reference: V5D (2G-60)
Brand: FRIGGA
Reference: P3110E
Brand: COMET
Reference: RXW-T11- 868
Brand: ONSET
Reference: MQ-210X
Brand: Apogee Instruments
Reference: SpectraPen SP 110
Brand: Photon System Instruments (PSI)
Reference: PX3PLN02S
Brand: VERIS
Reference: PX3DLX01S
Brand: VERIS
check_circle
check_circle